AM1620/1640/1660 series precision industrial PRTs are top choices when price-to-performance is considered. They cover a wide range of temperature from -200 °C to 670 °C with amazing accuracy of ±0.035°C at 0 °C, short term stability of ±0.01°C and fast respond time of 5 seconds. These industrial PRTs come with standard length 12-inch but customized dimensions are available per request.
To reach the best performance in stability and repeatability, the wire-wound sensing elements are specially designed to protect the platinum sensing wire from contamination at high temperature. A unique support structure and filling material provide the best balance among the hysteresis effect, mechanical shock and thermal shock performance. All of these probes conform to the standard 385 curve so the resistance ratio of the PRT follow DIN/IEC-751 curve precisely.
Features
Temperature range: -200 °C to 670 °C
Accuracy: ±0.035 °C at 0 °C
Long term drift: ±0.04 °C
Short term stability: 0.01 °C
Durable and shock resistance
Temperature Coefficient 0.00385
Follow DIN/IEC-751 precisely
Inconeltm sheath
Quick response time
Customized dimensions available
- SPECIFICATIONSTemperature Range1660: -200°C to 670°C1640: -200°C to 420°C1620: -60°C to 300°CResistance at 0 °CNominal 100 ΩTemperature Coefficient0.00385 Ω/ Ω/°CAccuracy±0.04°C at -200°C±0.035°C at 0°C±0.05°C at 200°C±0.09°C at 420°C±0.15°C at 660°CDrift±0.04°C at 0 °C after 100 hours at 420 °CShort Term Stability±0.01°CThermal Shock±0.007°C after 10 times thermal cycles from minimum to maximum temperaturesHysteresis<=0.01°CSelf-heating50 mW/°CResponse Time5 seconds for 63% response to step change in water moving at 3 feet per secondMeasurement Current0.5 mA or 1 mASensor Length32 mmSensor Location5 mm from tipInsulation Resistance>1000 MΩ at room temperatureSheath Material1660/1640: Inconeltm1620: 316 Stainless SteelDimension0.25 inch X 12 inch (6.35 mm X 305 mm)External LeadsTeflontm –insulated copper wire, 4 leads, 2.5 metersHandle Dimension15mm (OD) X 65 mm (L)Handle Temperature Range*-50°C to 180°CCalibration Options1660-12-T, PRT with NIST traceable calibration and data1640-12-T, PRT with NIST traceable calibration and data1620-12-T, PRT with NIST traceable calibration and dataNVLAP – Accredited Calibrations available per request
- Optional AccessoriesModelDescription9001Carrying Case
- Download