Features
Low volume design -only 4 liters of fluid required
Dual concentric chambers design provides the best temperature gradient uniformity in 300 mm immersion depth calibration bath.
Patent pending internal fluid agitator– internally liquid immersed rotor eliminating any leaks possibilities.
All 304 Stainless Steel construction – Internal chambers and enclosure.
No external cooling required for high temperature operation.
Small Exposed area of chamber fluid.
Externally heated fluid chamber assembly.
- MODEL 777 SPECIFICATIONS
Parameter Specification
Temperature range -80°C to 100°C; -40°C to 120°C; 40°C to 300°C
Stability ± 0.025°C at 100 °C
±0.03°C at 250°C
Uniformity 0.003°C using silicon oil
Heating Time 30 minutes (from 25°C to 100°C)
Cooling Time Less than 60 minutes (from 100°C to 25°C)
Stabilization Less than 45 minutes
Immersion Depth 406 mm
Display Resolution 0.01°C
Liquid Volume 4 liters
Dimensions 12 in x 20 in. x 32 ½ in. (304 mm x 508 mm x 825 mm)
Weight 30 kg
Heater Power 750 Watts
Consumption 6.5 amps at 120 VAC 60 Hz / 12 amps at 120VAC (low temperature)
Power Line Voltage 120 VAC 60Hz ±10%
Welded Parts Construction 316 Stainless Steel
Cabinet Construction Brushed or Polished 304 SS
- MODEL 777 Graphics
- Catalog